Ion fan tester

Jul 18, 2023 Leave a message

Ion fan tester

FMX-004 field tester

FMX-004 SIMCO

FMX-004 tester 2

I recently checked our ionization bar with two different ionizer testers (one portable and one with a regular 6*6 charging pad) and the portable charging pad was around 6V when taking readings over the same area and distance , the conventional charging board is -8V. And when I need to adjust the balance of the ion wind rod, which one should I trust?

Teacher reply: First, check the time of the last calibration of the two ion fan testers to ensure that they still meet the requirements and within the qualified time limit; second, check the measurement accuracy specifications of the equipment manufacturer; third, test the panel area Different results will also vary. The ion bar has multiple emitters arranged laterally along its axis, and it must be detected at the same distance and the same area covered by the ion wind during detection. If the emitter is DC, then the polarity of the ion wind is alternating, and the measurement results may be affected by the positioning of the plates.