Resistance measurement of anti-static work surface, floor, chair and wrist strap
A1 Measurement preparation
The object to be measured and the electrodes used for measurement must be clean. Before measurement, they should be cleaned twice with a volatile cleaning agent. After cleaning, wipe the surface with a clean low-cotton cloth and air-dry for at least 15 minutes to make the surface dry and clean.
A2 Test instrument requirements




The resistance value measured by the instrument must be at least 10 times higher and 10 times lower than the resistance value to be measured. When the resistance value to be measured is less than 1×105Ω, the instrument should be able to provide an open-circuit DC voltage of 10V±1 V. When the resistance value to be measured is in the range of 1X 10(5)Ω-1×10(11)Ω, it should be able to provide an open-circuit DC voltage of 100V±10V. The other two measuring wires should be relatively insulated.
A3 Electrode requirements
Each cylindrical electrode weighs 2.5Kg±0.5kg and has a diameter of 63.5mm±2.5mm. Each electrode has a contact of an electrical conductor material with a Shore A grade (International Rubber Hardness Scale) hardness of 50 to 70.
A4 Measurement method
A4.1 Measurement of ground resistance is to measure the resistance between a point on the surface and a ground connection point or an electrostatic grounding device.
A4.2 Measurement of surface resistance is to measure the resistance between two points on the surface.
A4.3 First, mark one or more measurement positions for the electrodes on the object to be measured, and use one or more measurement positions to represent the points on the surface that are grounded under use.
A4.4 Place the electrode at the marked measurement position, adjust the test instrument to the appropriate output voltage, and measure and record the resistance value after power-on for 30 seconds.
A4.5 The ambient temperature and humidity of all measurements must be recorded.
A4.6 Measurement of ground resistance after the working surface is installed is shown in Figure A1. Surface resistance measurement is shown in Figure A2.
A4.7 Measurement of ground resistance after the ground is installed is shown in Figure A3. Surface resistance measurement is shown in Figure A4. It is required that the ground resistance measurement and surface resistance measurement of each machine room floor be carried out at least 5 times, with 5 points randomly selected each time.
A4.8 The ground resistance measurement of the chair is shown in Figure A5. The surface resistance measurement is shown in Figure A6. The ground resistance measurement of the chair requires a piece of conductive sheet material as the floor contact point. Its volume resistivity should be less than 500n·cm, the hardness should be between 50 and 70 Shore A hardness, and the contact area should be not less than 254mmX254mm. The wheels or feet of the chair are placed on it.
Note: When the electrode must be supported by the operator or conductive material, the external non-contact surface of the electrode for measurement on the back of the chair should be covered with insulating material.
A4.9 The ground resistance measurement of the wristband is shown in Figure A7. The measuring instrument used is required to have a visible or audible display signal for the measured resistance value. The electrical connection of the wrist strap is achieved using a suitable connector for the plug and a rust-free iron hand contact plate with a size of at least 25.4mmX25.4mm. The operator wears the wrist strap and inserts the free end into the test instrument. The hand contact plate must be pressed until the equipment generates induction.