How to test the electrostatic performance of fabrics with anti-static work clothes
The anti-static work clothes tester is suitable for measuring the electrostatic performance of various types of fabrics such as fibers, yarns, or various plate-shaped products. The main body of the instrument consists of a corona discharge device and a probe detector. Using a given high-voltage electric field, the fabric is discharged for a fixed time to make the fabric induced by static electricity, thereby testing the size of the static electricity, the half-life of the static voltage decay, and the amount of static electricity residue to display the electrostatic performance of the fabric being tested.
1. Electrostatic potentiometer voltmeter; 2. Pressurization time; 3. Decay time; 4. Timing and constant pressure selection switch; 5. Clear key; 6. High voltage on; 7. Motor on; 8. Motor off; 9. High voltage off; 10. Main power switch; 11. High voltage adjustment knob; 12. High voltage indicator; 13. High voltage corona discharge head; 14. Protective device; 15. Sample turntable; 16. Electrostatic test head
III. Main technical features and process parameters
1. Power supply voltage 220V/50HZ
2. Measurement range of electrostatic voltage value 0~10KV
3. Half-life time limit range 0~99.99 hours, error ±0.01 seconds
4. Turntable speed 1500 rpm
5. Discharge time 0~99.99 hours adjustable (test set at 30 seconds)
6. Discharge distance between needle electrode and sample 20mm
7. Measurement distance between test probe and sample 15mm
8. Sample size: 45mm×60mm Three pieces
9. The instrument has been calibrated for static voltage measurement before it is issued
10. This instrument can be tested by timing method and constant pressure method respectively




IV. Operation instructions
1. Test preparation
The test is required to be carried out in an atmosphere with a relative humidity of 30% to 40% and a temperature of 20±2℃. After the sample is pre-dried at a temperature of 50℃ for 30 minutes, it is humidified and balanced under the test atmospheric conditions for 5 hours. Plug in the power socket, turn on the main power switch, and rotate the high voltage adjustment knob to make the high voltage display show 10KV. Turn on the power switch of the control box and press the "clear" button to make the attenuation voltage on the static meter display table show zero. The preset pressurization time is 30 seconds on the "pressurization time" timer.
The anti-static work clothes detector is suitable for measuring the electrostatic properties of various types of fabrics such as fibers and yarns or various plate-shaped products. The main instrument consists of a corona discharge device and a probe detector. By using a given high-voltage electric field to discharge the fabric for a fixed time, the fabric will be induced to induce static electricity, so that the static electricity quantity, the half-life of static voltage decay, and the static electricity residual amount can be tested to show the static electricity performance of the fabric being tested.