Test Considerations for Ion Static Elimination Generator
1) During the test, there shall be no metal objects, electrical equipment, personnel and other electrostatically charged bodies within 100cm of the phase distance between the test instrument and the ion static eliminator;
2) The airflow disturbance in the horizontal and vertical layers generated by the centralized air conditioner, fan or other air conditioning device in the test room;
3) The relative humidity of the test environment is (40±3)%;